Pathology Image Analysis (PIA)
Sequence of Operation
| Slide preparation, bar coding |
Entry of slides data |
Scanning | Review and regions selection |
Automated scoring and analysis | Export of results and images |
- Histological slides should be clean, ideally without staining and coverslipping artifacts such as air-bubbles, folds, and background staining.
- Tissue should be cut with a new blade, uneven tissue can cause focus failures
- Coverslipping: If the manual coverslipping technique is used, be sure not to use excessive mounting medium
- Coverslips should be cleaned with alcohol and lens cleaning paper just prior to running on the imaging system
- Barcodes are provided in convenient rolls. There are 1000 duplicate barcodes per roll. Two identical labels appear adjacent to each other on the roll.
- One barcode should be placed on each slide, on the right hand upper corner, with the number facing outward
- The laser reads barcode lines which are horizontal, therefore pre-existing labels on the slide which are vertical will not be detected by the internal laser scanner
- The additional barcode should be placed on the worksheet in the appropriate position
- Positive Controls: a known positive control may be placed on the same slide as the patient/sample tissue, and/or a known positive control may be placed on a separate slide.
- Negative Controls; an additional slice of tissue is cut from the patient/sample block and stained with an isotope antibody and/or a known negative tissue may be used.

The Slide Carriers
- 4 Slides can be placed in a slide carrier, 25 carriers can be stacked in the input hopper
- Check the carriers before use to ensure they are intact, not warped or broken in any way
- Slides should be placed in the carrier under the bottom tab, push down on the spring and secure firmly under the top tab
- The slide carrier should be on top of the wheel in the input hopper
- Do not label the slide carriers with pen, pencils or labels in the area above the slides









